气相分子吸收光谱法测定硫化物校准曲线的研究
Determination of Sulfide Calibration Curve by Using Gas-Phase Molecular Absorption Spectrometry
摘要:
本文主要阐述了气相分子吸收光谱法测定硫化物时对其校准曲线斜率、截距参考值的分析。通过收集实验室近三年采用气相分子吸收光谱法测定硫化物时制作校准曲线的一系列原始数据,并用数理统计方法对校准曲线及其有关指标进行了详细讨论,得出气相分子吸收光谱法测定硫化物校准曲线的斜率参考值范围为:0.0961 ± 0.0013,截距参考值范围:−0.0018 ± 0.0030。此结果对实验室分析人员判定硫化物项目校准曲线制作是否合格具有一定的参考价值。
Abstract:
This paper mainly analyzed the reference value of the slope rate and intercept from the sulfide’s calibration curve by using gas-phase molecular absorption spectrometry. The calibration curve and its related indexes were discussed in detail by using mathematical statistics method through nearly three-year original data in laboratories. Results showed that the reference range of slope of the gas-phase molecular absorption spectrometry calibration curve is 0.0961 ± 0.0013, and that of intercept is −0.0018 ± 0.0030. These results provide reference for determining eligibility criteria of sulfide calibration curve.
参考文献
[1]
|
陈秋红, 施建兵, 藏平安. 气相分子仪测定水中的硫化物[J]. 污染防治技术, 2006, 19(2): 65-66.
|
[2]
|
林志彬. 气相分子吸收光谱仪在水中硫化物检测的应用[J]. 中国新技术新产品, 2011(10): 160.
|
[3]
|
陈小霞. 气相分子吸收光谱法对水体中硫化物的测定应用[J] 海峡科学, 2014, 88(4): 33-34.
|
[4]
|
连正豪. 气相分子吸收光谱法测定水中硫化物浓度不确定度评定[J]. 北方环境, 2013, 29(5): 155-157.
|
[5]
|
国家环境保护总局科技标准司, 上海宝钢工业检测公司宝钢环境监测站, 苏州市环境监测中心站, 等 (2005) (HJ/T200─2005)水质硫化物的测定气相分子吸收光谱法[S].北京: 中国环境出版社, 2005: 1-3.
|
[6]
|
邓勃. 关于校准曲线建立和应用中一些问题的探讨[J]. 中国无机分析化学, 2011, 1(3): 2-3.
|
[7]
|
盛骤, 谢式千, 潘承毅. 概率论与数理统计[M]. 北京: 高等教育出版社, 2008: 385.
|