超声外科手术系统输出声功率的测量方法研究
Measurement of Acoustic Output Poweron Ultrasonic-Surgical Systems
DOI: 10.12677/APP.2017.712043, PDF,  被引量   
作者: 黄 涛*, 王龙军:国家食品药品监督管理局湖北医疗器械质量监督检验中心,湖北 武汉
关键词: 超声外科手术系统输出声功率水听器无损检测Ultrasonic-Surgical Systems Output Acoustic Power Hydrophone Nondestructive Detection
摘要: 该文综述讨论了测试超声外科手术系统输出声功率原理、方法以及重要性和必要性,并采用水听器法对超声止血刀系统输出声功率进行了实验测量,并对实验数据进行了初步分析讨论。该测量方法与实验结果对评价超声外科手术系统的安全性、有效性有重要指导意义。
Abstract: This paper reviews the principle, method, importance and necessity of testing the output power of ultrasonic surgical system, and uses the hydrophone method to measure the output sound power of the ultrasonic hemostatic knife system. The experimental data are analyzed and discussed. The measurement method and experimental results are of great significance to evaluate the safety and efficacy of ultrasound surgical system.
文章引用:黄涛, 王龙军. 超声外科手术系统输出声功率的测量方法研究[J]. 应用物理, 2017, 7(12): 351-357. https://doi.org/10.12677/APP.2017.712043

参考文献

[1] 尚志远. 检测声学原理及应用[M] 冯若, 王智彪. 实用超声治疗学[M]. 北京: 科学技术文献出版社, 2002.
[2] Takahashi, H. and Murakami, E. (2010) Ultrasonic Surgical System. Digestive Endoscopy, 14, 163-166.
[3] EPL/87 (1998) Ultrasonics-Surgical Systems-Measurement and Declaration of the Basic Output Characteristics. BS EN 61847:1998, IEC 61847:1998.
[4] . 西安: 西北大学出版社, 1996.
[5] 国家武汉医用超声波仪器质量监督检测中心. YY/T 0644-2008超声外科手术系统基本输出特性的测量和公布[S]. 北京: 中国国家标准化管理委员会, 2008.
[6] Petošić, A.I., Jambrošić, B., et al. (2007) Measuring Output Acoustic Power of an Ultrasonic Surgical System Using the Calorimetric Method; Proceedings of the International Congress on Acoustics Official Publication. 19th International Congress on Acoustics, Official Publication.
[7] ONDA (2015) Hydrophone Handbook Datasheet. ONDA, USA, 33.
[8] 上海交通大学, 等. GB/T 19890-2005声学 高强度聚焦超声(HIFU)声功率和声场特性的测量[S]. 北京: 中国国家标准化管理委员会, 2005.
[9] Kollmann, C., Haar, G.T., Doležal, L., et al. (2005) Ultrasound, Output, Safety, Thermal Index, Mechanical Index. Journal of the American Society of Echocardiography, 9, 216-220.