BD流式细胞仪原理与典型故障分析
BD Flow Cytometer Principle and Typical Malfunction Analysis
DOI: 10.12677/IaE.2018.63015, PDF,   
作者: 丘秀生*, 曾彩辉, 张丽婷:中山大学附属第三医院,广东 广州
关键词: 流式细胞仪原理故障分析Flow Cytometer Principle Malfunction Analysis
摘要: 流式细胞仪是对细胞进行自动分析和分选的装置,已经渐渐成为生物科学重要的分析仪器。本文介绍了BD流式细胞仪的原理,并对流式细胞仪常见的典型故障做了阐述,进行分析后,并对故障的排除过程进行了详细的阐述。流式细胞仪维中设备故障可分为液路故障、硬件故障、软件故障和三者兼具的情况,只有在使用过程中不断积累,才能快速判断故障,并迅速的排除。
Abstract: Flow cytometry is a device for automatic cell analysis and cell separation. In this paper, the princi-ple of BD flow cytometry is introduced, and the common faults of flow cytometry are described. After analysis, the troubleshooting process is described in detail. Flow cytometry instrument equipment failure can be divided into liquid road faults, hardware failure, software failure, and all three cases. Only by accumulating in use can we fast judge the fault, and quickly solve the problem.
文章引用:丘秀生, 曾彩辉, 张丽婷. BD流式细胞仪原理与典型故障分析[J]. 仪器与设备, 2018, 6(3): 95-98. https://doi.org/10.12677/IaE.2018.63015

参考文献

[1] 李超, 韩金路, 王玉刚, 孙英凯. 流式细胞仪的工作原理及应用[J]. 中国实用医药, 2009, 4(20): 235-236.
[2] 李培援, 唐国山. XL系列流式细胞仪的原理及常见故障维修[J]. 医疗设备信息, 2006, 21(1): 86.
[3] 张艺. 流式细胞仪构成与工作原理[J]. 医疗设备信息, 2005, 20(8): 25-26.
[4] 董荔红. CLOUTER流式细胞仪几类常见故障的排除和预防措施[J]. 现代检验医学杂志, 2008(1): 72-73.