SAR测量中系统验证的测量与分析
Measurement and Analysis of System Verification in SAR Measurement
DOI: 10.12677/JA.2023.122003, PDF,   
作者: 李 晋:河北科技大学国家无线电监测中心检测中心,北京;安 晴:中国传媒大学信息与通信工程学院,北京;李 彬:北京交通大学国家无线电监测中心检测中心,北京
关键词: 电磁能量吸收比值(SAR)电磁安全系统验证Specific Absorption Ratio (SAR) Electromagnetic Safety System Validation
摘要: 随着手机等含有射频发射天线的物品在人们生活中逐渐普及,在设计过程中就必须考虑到电磁能量吸收比值(SAR)。该指标的大小能够衡量射频发射器不会对人体产生危害,保证电磁安全,因此是在制作发射器时非常重要的因素。为了保证SAR测量的准确性,在测量系统软件被更改升级,或硬件设备需要调整时,需要及时进行SAR系统验证。本文通过六个步骤进行SAR系统验证,测量结果均符合测试规定中的指标,能够验证完整测量系统的精度符合要求。
Abstract: With the increasing popularity of items containing radio frequency transmitting antennas such as mobile phones, the specific absorption rate (SAR) must be considered in the design process. The size of this indicator can measure that the RF transmitter will not cause harm to human body and ensure electromagnetic safety, so it is a very important factor when making the transmitter. In or-der to ensure the accuracy of SAR measurement, when the measurement system software is changed or upgraded, or the hardware equipment needs to be adjusted, SAR system verification needs to be carried out in a timely manner. In this paper, SAR system verification is carried out in six steps, and the measurement results are in line with the indicators specified in the test, which can verify that the accuracy of the complete measurement system meets the requirements.
文章引用:李晋, 安晴, 李彬. SAR测量中系统验证的测量与分析[J]. 天线学报, 2023, 12(2): 19-27. https://doi.org/10.12677/JA.2023.122003

参考文献

[1] Lu, M. and Ueno, S. (2012) Comparison of Specific Absorption Rate Induced in Brain Tissues of a Child and an Adult Using Mobile Phone. Journal of Applied Physics, 111, 07B3111-07B3113. [Google Scholar] [CrossRef
[2] Iqbal-Faruque, M.R., Aisyah-Husni, N., Ikbal-Hossain, M. et al. (2014) Effects of Mobile Phone Radiation onto Human Head with Variation of Holding Cheek and Tilt Positions. Journal of Applied Research & Technology, 12, 871-876. [Google Scholar] [CrossRef
[3] Bahramzy, P., Svendsen, S., Jagielski O. and Pedersen, G.F. (2015) SAR Study of Mobile Phones as a Function of Antenna Q. IEEE Transaction on Antennas and Propagation, 63, 4139-4147. [Google Scholar] [CrossRef
[4] Zhang, H.H., Yu, G.G., Liu, Y., Fang, Y.X., Shi, G. and Wang, S. (2021) Design of Low-SAR Mobile Phone Antenna: Theory and Applications. IEEE Transaction on Antennas and Propagation, 69, 698-707. [Google Scholar] [CrossRef
[5] Le, D.T., Hamada, L., Watanabe, S. and Onishi, T. (2017) A Fast Estimation Technique for Evaluating the Specific Absorption Rate of Multiple-Antenna Transmitting Device. IEEE Transaction on Antennas and Propagation, 65, 1947-1957. [Google Scholar] [CrossRef
[6] Li, H. Tsiaras, A. and Lau, B.K. (2017) Analysis and Estimation of MIMO-SAR for Multi-Antenna Mobile Handsets. IEEE Transaction on Antennas and Propagation, 65, 1522-1527. [Google Scholar] [CrossRef
[7] Zhao, K., Zhang, S., Ying, Z., Bolin, T. and He, S. (2013) SAR Study of Different MIMO Antenna Designs for LTE Applica-tion in Smart Mobile Handsets. IEEE Transaction on Antennas and Propagation, 61, 3270-3279. [Google Scholar] [CrossRef
[8] Wu, T., Zhou, X., et al. (2020) Proficiency Testing for Complex Permittivity Measurements of Tissue Equivalent Liquid Used in SAR Assessment. IEEE Access, 8, 210592-210596. [Google Scholar] [CrossRef