手术室温度对FS-LASIK术后结果的影响
The Effect of Operating Room Temperature on FS-LASIK Outcomes
摘要: 目的:探讨手术室温度对飞秒激光制瓣的准分子激光原位角膜磨镶术(Femtosecond laser-assisted in situ keratomileusis, FS-LASIK)术后结果的影响。方法:回顾性研究。收集我院2023-05/2023-10行FS-LASIK的病例,抽取年龄、屈光度数、角膜厚度、角膜曲率、手术室湿度差异小,但手术时温度不同的病例分为三组,每组20人40只眼,共60人120只眼。A组手术室温度为20.0℃~20.5℃,B组为22.0℃~22.5℃,C组为24.0℃~24.5℃。手术由同一位手术医生、同样的设备完成,术后统一用药。比较三组术后3个月时裸眼视力与屈光度的差异情况。结果:不同手术温度组在术后3月比较时,未发现裸眼视力及屈光度之间存在统计学差异(P > 0.05)。三组患者术后3个月裸眼视力、等效球镜度数不满足正态分布,采用Kruskal-Wallis法进行检验。术后裸眼视力比较无统计学意义(H = 2.648, P = 0.266)。三组患者术后等效球镜度数比较无统计学意义(H = 1.010, P = 0.604)。结论:空气中的温度及湿度的明显变化才能对准分子激光能量产生显著影响。湿度可能较温度的影响更大。本文研究结果未发现在设备商推荐温度范围内变化时,对FS-LASIK术后视力屈光度产生有统计学意义的显著影响。
Abstract: Aim: To explore the effect of operating room temperature on the postoperative outcomes of femtosecond laser-assisted in situ keratomileusis (FS-LASIK). Methods: Retrospective study. Collecting cases that underwent FS-LASIK in our hospital from May 2023 to October 2023, cases were extracted with small differences in age, diopters, corneal thickness, corneal curvature, and operating room humidity. Cases with different temperatures during surgery were divided into three groups. Each group has 20 people with 40 eyes, a total of 60 people with 120 eyes. The operating room temperature in group A was 20.0˚C~20.5˚C, in group B was 22.0˚C~22.5˚C, and that in group C was 24.0˚C~24.5˚C. The surgery was completed by the same surgeon and the same equipment, and the same postoperative medication was used. The differences in naked eye visual acuity and diopters between the three groups were compared 3 months after surgery. Results: Comparing different surgical temperature groups 3 months after surgery, no statistical difference was found between naked eye visual acuity and diopters (P > 0.05). The naked visual acuity and spherical equivalent refraction of the three groups 3 months after surgery did not satisfy the normal distribution, and the Kruskal-Wallis method was used to test. There was no statistical significance in postoperative naked eye visual acuity (H = 2.648, P = 0.266). There was no statistical significance in the postoperative spherical equivalent refraction of the three groups (H = 1.010, P = 0.604). Conclusion: Only significant changes in temperature and humidity in the air can have a significant impact on excimer laser energy. Humidity may have a greater impact than temperature. The results of this study did not find that changes in the temperature range recommended by the equipment manufacturer have a statistically significant impact on the visual acuity and diopters after FS-LASIK.
文章引用:章文成, 何晶晶, 戎志銮. 手术室温度对FS-LASIK术后结果的影响[J]. 临床医学进展, 2024, 14(7): 990-995. https://doi.org/10.12677/acm.2024.1472105

参考文献

[1] Hood, C., Shtein, R., Veldheer, D., Hussain, M., Niziol, L., Musch, D., et al. (2016) The Effect of Humidity and Temperature on Visual Outcomes after Myopic Corneal Laser Refractive Surgery. Clinical Ophthalmology, 10, 2231-2236. [Google Scholar] [CrossRef] [PubMed]
[2] Albietz, J.M., McLennan, S.G. and Lenton, L.M. (2003) Ocular Surface Management of Photorefractive Keratectomy and Laser in Situ Keratomileusis. Journal of Refractive Surgery, 19, 636-644. [Google Scholar] [CrossRef] [PubMed]
[3] Shtein, R.M., Michelotti, M.M., Kaplan, A. and Mian, S.I. (2012) Association of Surgeon Experience with Outcomes of Femtosecond LASIK. Ophthalmic Surgery, Lasers and Imaging Retina, 43, 489-494. [Google Scholar] [CrossRef] [PubMed]
[4] Benabidi, S., Frings, A., Druchkiv, V. and Katz, T. (2024) Influence of the Patient’s Age on the Safety, Efficacy, and Prediction Accuracy of the Microkeratome in Laser-Assisted in Situ Keratomileusis. Scientific Reports, 14, Article No. 1972. [Google Scholar] [CrossRef] [PubMed]
[5] Mohamed Mostafa, E. (2015) Effect of Flat Cornea on Visual Outcome after Lasik. Journal of Ophthalmology, 2015, Article 794854. [Google Scholar] [CrossRef] [PubMed]
[6] Gomel, N., Negari, S., Frucht-Pery, J., Wajnsztajn, D., Strassman, E. and Solomon, A. (2018) Predictive Factors for Efficacy and Safety in Refractive Surgery for Myopia. PLOS ONE, 13, e0208608. [Google Scholar] [CrossRef] [PubMed]
[7] Zhang, M., Li, X. and Hu, Q. (2022) Analysis on the Correlation between Long-term Refractive Regression and Visual Quality after FS-LASIK. Current Eye Research, 47, 824-831.
[8] Chlasta-Twardzik, E., Górecka-Nitoń, A., Nowińska, A. and Wylęgała, E. (2021) The Influence of Work Environment Factors on the Ocular Surface in a One-Year Follow-up Prospective Clinical Study. Diagnostics, 11, Article 392. [Google Scholar] [CrossRef] [PubMed]
[9] Wallace, D.A. (2014) Reporting Acuity Outcomes and Refractive Accuracy after Lasik. Journal of Refractive Surgery, 30, 798-798. [Google Scholar] [CrossRef] [PubMed]
[10] Vinh, L.M., Mai, P.H., Nguyễn, Đ., et al. (2022) Comparison of Clinical Outcomes between Wavelight Allegretto Eye-Q and Schwind Amaris 1050RS in High Myopic Astigmatism. Science and Technology Development Journal: Health Sciences, 3, 427-435. [Google Scholar] [CrossRef
[11] 田艳明, 鞠燕, 王文强, 等. 西部高温干燥地区LASIK治疗近视过矫影响因素分析及对策[J]. 眼科新进展, 2006, 26(11): 851-852.
[12] 昌宏发, 白宗禧, 曲兴雷, 等. 西藏高原地区LASIK治疗近视过矫的临床分析[J]. 国际眼科杂志, 2015, 15(2): 311-313.
[13] 李爱萍. 准分子激光系统的稳定性与温度之间的关系[J]. 医疗装备, 2002, 15(2): 25-26.
[14] Schuh, A., Kolb, C.M., Mayer, W.J., Vounotrypidis, E., Kreutzer, T., Kohnen, T., et al. (2021) Comparison of Changes in Corneal Volume and Corneal Thickness after Myopia Correction between LASIK and Smile. PLOS ONE, 16, e0250700. [Google Scholar] [CrossRef] [PubMed]
[15] Schena, E., Silvestri, S., Franzesi, G.T., Cupo, G., Carito, P. and Ghinelli, E. (2006) Theoretical Model and Design of a Device to Reduce the Influence of Environmental Factors on Refractive Surgery Outcomes. 2006 International Conference of the IEEE Engineering in Medicine and Biology Society, New York, 30 August-3 September 2006, 343-346. [Google Scholar] [CrossRef] [PubMed]
[16] Mimouni, M., Vainer, I., Shapira, Y., Levartovsky, S., Sela, T., Munzer, G., et al. (2016) Factors Predicting the Need for Retreatment after Laser Refractive Surgery. Cornea, 35, 607-612. [Google Scholar] [CrossRef] [PubMed]
[17] Seider, M.I., McLeod, S.D., Porco, T.C. and Schallhorn, S.C. (2013) The Effect of Procedure Room Temperature and Humidity on LASIK Outcomes. Ophthalmology, 120, 2204-2208. [Google Scholar] [CrossRef] [PubMed]
[18] Randleman, J.B., White, A.J., Lynn, M.J., Hu, M.H. and Stulting, R.D. (2009) Incidence, Outcomes, and Risk Factors for Retreatment after Wavefront-Optimized Ablations with PRK and Lasik. Journal of Refractive Surgery, 25, 273-276. [Google Scholar] [CrossRef] [PubMed]
[19] Verma, S., Kehrer, T., Hesser, J. and Arba Mosquera, S. (2019) Analysis of Impact of Humidity and Temperature on Excimer Laser Ablation of Polyethylene Terephthalate, Polymethylmethacrylate, and Porcine Corneal Tissue. Lasers in Surgery and Medicine, 52, 627-638. [Google Scholar] [CrossRef] [PubMed]
[20] Walter, K.A. and Stevenson, A.W. (2004) Effect of Environmental Factors on Myopic LASIK Enhancement Rates. Journal of Cataract and Refractive Surgery, 30, 798-803. [Google Scholar] [CrossRef] [PubMed]
[21] de Ortueta, D., von Rüden, D., Magnago, T. and Mosquera, S.A. (2014) Influence of Stromal Refractive Index and Hydration on Corneal Laser Refractive Surgery. Journal of Cataract and Refractive Surgery, 40, 897-904. [Google Scholar] [CrossRef] [PubMed]
[22] Kim, W. and Jo, J. (2001) Corneal Hydration Affects Ablation during Laser in Situ Keratomileusis Surgery. Cornea, 20, 394-397. [Google Scholar] [CrossRef] [PubMed]
[23] Frings, A., Neuhaus-Richard, I., Görsch, I., Druchkiv, V., Katz, T., Linke, S., et al. (2015) Does Outside Environmental Humidity Influence the Outcome of Laser Refractive Surgery? Results from the Hamburg Weather Study. Journal of Clinical Ophthalmology and Research, 3, 133-138. [Google Scholar] [CrossRef
[24] Chatterjee, A. and Shah, S. (1997) Seasonal Variations in Refractive Results Following Excimer Laser Photorefractive Keratectomy. Journal of Refractive Surgery, 13, S447-S449. [Google Scholar] [CrossRef] [PubMed]
[25] de Souza, I.R.U., de Queiroz Urbano de Souza, A.P., de Queiroz Urbano de Souza, A.P., Figueiredo, P., Jesus, R.S. and Kara-José, N. (2001) Influence of Temperature and Humidity on Laser in Situ Keratomileusis Outcomes. Journal of Refractive Surgery, 17, S202-S204. [Google Scholar] [CrossRef] [PubMed]
[26] Wernli, J., Schumacher, S., Wuellner, C., Donitzky, C. and Mrochen, M. (2012) Initial Surface Temperature of PMMA Plates Used for Daily Laser Calibration Affects the Predictability of Corneal Refractive Surgery. Journal of Refractive Surgery, 28, 639-644. [Google Scholar] [CrossRef] [PubMed]
[27] Fisher, B.T. and Hahn, D.W. (2007) Development and Numerical Solution of a Mechanistic Model for Corneal Tissue Ablation with the 193 nm Argon Fluoride Excimer Laser. Journal of the Optical Society of America A, 24, 265-277. [Google Scholar] [CrossRef] [PubMed]