IgM-κ型M蛋白对凝血试验的干扰一例及原因分析
Interference of IgM-κ Type M Protein on Coagulation Test: A Case Report and Analysis of Its Cause
DOI: 10.12677/acm.2025.1541103, PDF,   
作者: 朱瑞玉, 彭湘明*:暨南大学附属广州红十字会医院检验科,广东 广州
关键词: M蛋白干扰凝血检测凝固曲线M Protein Interference Blood Coagulation Test Curve of Coagulation
摘要: 目的:通过一例华氏巨球蛋白血症患者探讨M蛋白对凝血试验的干扰,提高检验者对M蛋白的认识,避免影响临床的诊治。方法:查看凝固曲线发现干扰因素,通过进一步实验室检查明确干扰物质并分析原因。结果:大量IgM-κ型M蛋白在溶液环境改变时易出现沉淀干扰凝血实验的检测。结论:M蛋白可通过溶液浊度的改变干扰凝血实验的检测,当发现患者凝血结果异常或与临床表现不符时,应及时与临床沟通,并查看凝固曲线,必要时手工法进行复查。
Abstract: Objective: To explore the interference of M protein on coagulation test through a case of Waldenstrom macroglobulinemia, so as to improve the understanding of M protein in the laboratory and avoid affecting clinical diagnosis and treatment. Methods: The interfering factors were found by checking the coagulation curve, and the interfering substances were identified by further laboratory tests and the reasons were analyzed. Results: A large amount of IgM-κ type M protein was easy to precipitate and interfere with the detection of coagulation test when the solution environment changed. Conclusions: M protein can interfere with the detection of coagulation test by changing the turbidity of the solution. When the patient’s coagulation results are found to be abnormal or inconsistent with the clinical manifestations, we should communicate with the clinic in time and check the coagulation curve, and manually recheck if necessary.
文章引用:朱瑞玉, 彭湘明. IgM-κ型M蛋白对凝血试验的干扰一例及原因分析[J]. 临床医学进展, 2025, 15(4): 1649-1653. https://doi.org/10.12677/acm.2025.1541103

参考文献

[1] Cook, L. and Macdonald, D.H.C. (2007) Management of Paraproteinaemia. Postgraduate Medical Journal, 83, 217-223. [Google Scholar] [CrossRef] [PubMed]
[2] Sarkar, R. (2024) Paraprotein Interferences: Insights from a Short Study Involving Multiple Platforms and Multiple Measurands. eJIFCC, 35, 120-131.
[3] 周伟, 安倍莹, 刘睿博, 等. 常见内源性干扰物对凝血试验结果的影响分析[J]. 临床检验杂志, 2023, 41(1): 38-43.
[4] Swerdlow, S.H., Campo, E., Pileri, S.A., Harris, N.L., Stein, H., Siebert, R., et al. (2016) The 2016 Revision of the World Health Organization Classification of Lymphoid Neoplasms. Blood, 127, 2375-2390. [Google Scholar] [CrossRef] [PubMed]
[5] Sloop, G.D., Moore, C., Pop, G., Weidman, J.J. and St. Cyr, J.A. (2023) New Onset Anemia, Worsened Plasma Creatinine Concentration, and Hyperviscosity in a Patient with a Monoclonal IGM Paraprotein. Cureus, 15, e41657. [Google Scholar] [CrossRef] [PubMed]
[6] King, R.I. and Florkowski, C.M. (2010) How Paraproteins Can Affect Laboratory Assays: Spurious Results and Biological Effects. Pathology, 42, 397-401. [Google Scholar] [CrossRef] [PubMed]
[7] 乔建启, 郭丽, 褚娜利, 等. 淋巴浆细胞淋巴瘤(华氏巨球蛋白血症)伴血钙假性增高1例[J]. 中国实验诊断学, 2022, 26(10): 1553-1555.
[8] Ting, H.Y., Ting, I.P.L., Lo, S.C. and Loh, T.P. (2022) Falsely Low Serum Creatinine Caused by Immunoglobulin M Paraprotein Interference with Enzymatic Method. Pathology, 54, 959-962. [Google Scholar] [CrossRef] [PubMed]
[9] Song, L., Tong, K.H. and Chin, C.D. (2019) Gelation of Monoclonal Protein Was the Cause of Interference with a Total Bilirubin Assay. The Journal of Applied Laboratory Medicine, 3, 1054-1058. [Google Scholar] [CrossRef] [PubMed]
[10] Tichy, M., Friedecky, B., Budina, M., Maisnar, V., Buchler, T., Holeckova, M., et al. (2009) Interference of IGM-Λ Paraprotein with Biuret-Type Assay for Total Serum Protein Quantification. Clinical Chemistry and Laboratory Medicine, 47, 235-236. [Google Scholar] [CrossRef] [PubMed]
[11] Margetic, S., Celap, I., Dukic, L., Vukasovic, I. and Virovic-Jukic, L. (2016) Interference of M-Protein on Prothrombin Time Test—Case Report. Biochemia Medica, 26, 248-254. [Google Scholar] [CrossRef] [PubMed]
[12] Alberti, M.O., Drake, T.A. and Song, L. (2015) The pH of Chemistry Assays Plays an Important Role in Monoclonal Immunoglobulin Interferences. Practical Laboratory Medicine, 3, 8-16. [Google Scholar] [CrossRef] [PubMed]
[13] 付帅, 徐丝, 杨敏, 等. 血清胱抑素C公式和肌酐公式对评估中国人肾小球滤过率精准度的比较[J]. 临床肾脏病杂志, 2020, 20(5): 351-356+384.
[14] Mase, H., Hamano, N., Mizuhara, R., Nozaki, T., Saso, T., Wada, T., et al. (2020) Falsely Elevated Serum Creatinine Associated with IGM Paraproteinemia. Kidney International Reports, 5, 377-381. [Google Scholar] [CrossRef] [PubMed]