1,4-丁二醇中的39种金属元素的检测方法
Method for Detecting 39 Metals in 1,4-Butanediol
摘要: 建立了一种快速测定1,4-丁二醇中39种痕量金属元素的检测方法。使用电感耦合等离子体质谱检测有机物时,加入氧气,减少积碳,避免熄火。雾化室温度需设定为2℃,提供稳定的进样环境。1,4-丁二醇,黏稠且凝固点为20.1℃,与N-甲基吡咯烷酮混合后,混合物可以直接进样检测分析,从而减少预处理的检测误差。加标回收率在92.21%~105.96%之间,验证了方法的准确性。此方法检测1,4-丁二醇中的金属元素,具有高效、干扰小、污染少、灵敏度高的优点。
Abstract: A rapid detection method for 39 trace metal elements in 1,4-butanediol was established. When using inductively coupled plasma mass spectrometry (ICP-MS) to detect organic substances, oxygen is added to reduce carbon deposition and avoid flameout. The temperature of the nebulization chamber should be set at 2˚C to provide a stable sample introduction environment. 1,4-butanediol is viscous with a freezing point of 20.1˚C. After mixing it with N-methylpyrrolidone, the mixture can be directly injected for detection and analysis, thus reducing the detection errors caused by sample pretreatment. The recoveries of the spiked samples range from 92.21% to 105.96%, which verifies the accuracy of the method. This method for detecting metal elements in 1,4-butanediol has the advantages of high efficiency, less interference, low pollution, and high sensitivity.
文章引用:李清霞, 梁斌, 张勇娟, 梁亚娟, 胜继伟. 1,4-丁二醇中的39种金属元素的检测方法[J]. 材料科学, 2025, 15(7): 1443-1451. https://doi.org/10.12677/ms.2025.157153

参考文献

[1] 魏国玉, 侯艺琳. Reppe法1,4-丁二醇铜铋催化剂国产应用[J]. 上海化工, 2024, 49(4): 29-32.
[2] 吴慧玲, 孟令臣, 张思琦, 等. 1, 4-丁二醇制备γ-丁内酯脱氢工艺[J]. 石油化工, 2025, 54(3): 297-303.
[3] Raju, M.A., Gidyonu, P., Nagaiah, P., Rao, M.V., Raju, B.D. and Rao, K.S.R. (2019) Mesoporous Silica-Supported Copper Catalysts for Dehydrogenation of Biomass-Derived 1,4-Butanediol to Gamma Butyrolactone in a Continuous Process at Atmospheric Pressure. Biomass Conversion and Biorefinery, 9, 719-726. [Google Scholar] [CrossRef
[4] 杨虹. 原子吸收光谱法在测定水中重金属的应用研究[J]. 山西化工, 2021, 41(5): 74-77.
[5] 冯先进, 郑国经. 电感耦合等离子体发射光谱与质谱分析技术和应用研究进展[J]. 中国无机分析化学, 2025, 15(3): 353-362.
[6] Mdluli, N.S., Nomngongo, P.N. and Mketo, N. (2024) Ionic Liquid Assisted Extraction Induced by Emulsion Breaking for Extraction of Trace Metals in Diesel, Gasoline and Kerosene Prior to ICP-OES Analysis. Heliyon, 10, e26605. [Google Scholar] [CrossRef] [PubMed]
[7] Hirata, T., Kobayashi, K., Khoo, H.H., Shikino, O. and Asanuma, H. (2024) Detection of Several Volatile Organic Compounds through Ar+ Induced Chemical Ionisation Using Inductively Coupled Plasma-Tandem Mass Spectrometry (ICP-MS/MS). The Analyst, 149, 5174-5183. [Google Scholar] [CrossRef] [PubMed]
[8] 汤地生, 杨婉晴, 张宇, 等. 电感耦合等离子体质谱(ICP-MS)法测定高钨基体样品中稀土元素的含量[J]. 中国无机分析化学, 2024, 14(4): 457-463.
[9] 黄丹宇. 电感耦合等离子体质谱法测定镁合金中的痕量元素[J]. 分析测试技术与仪器, 2023, 29(3): 328-333.
[10] Lossow, K., Schlörmann, W., Tuchtenhagen, M., Schwarz, M., Schwerdtle, T. and Kipp, A.P. (2023) Measurement of Trace Elements in Murine Liver Tissue Samples: Comparison between ICP-MS/MS and TXRF. Journal of Trace Elements in Medicine and Biology, 78, Article 127167. [Google Scholar] [CrossRef] [PubMed]
[11] Sim, K.S., Kim, H., Hur, S.H., Na, T.W., Lee, J.H. and Kim, H.J. (2024) Geographical Origin Discriminatory Analysis of Onions: Chemometrics Methods Applied to ICP-OES and ICP-MS Analysis. Food Research International, 175, Article 113676. [Google Scholar] [CrossRef] [PubMed]
[12] 万曦, 田海花, 刘涌, 等. ICP-MS测定电子级丙二醇甲醚中的钠、镁、钙、铁和铅含量[J]. 2023, 42(1): 90-94.
[13] 张文媚, 吕玉平, 闻环, 等. 有机溶剂稀释-电感耦合等离子体发射质谱法测定原油砷含量[J]. 广东化工, 2022, 49(3): 179-181.
[14] 李丹. 基于有机溶剂前处理的ICP-MS法直接测定盐酸溴己新中的汞含量[J]. 广州化工, 2025, 53(8): 160-163.
[15] 张磊, 唐辉, 王辉, 等. 电感耦合等离子体质谱法直接测定有机相中痕量锆[J]. 质谱学报, 2015, 36(2): 156-160.
[16] 李春华, 陈鹰, 韩宝英, 等. 化学试剂 电感耦合等离子体质谱分析方法通则: GB/T 39486-2020 [S]. 2020.