1,4-丁二醇中的39种金属元素的检测方法
Method for Detecting 39 Metals in 1,4-Butanediol
摘要: 建立了一种快速测定1,4-丁二醇中39种痕量金属元素的检测方法。使用电感耦合等离子体质谱检测有机物时,加入氧气,减少积碳,避免熄火。雾化室温度需设定为2℃,提供稳定的进样环境。1,4-丁二醇,黏稠且凝固点为20.1℃,与N-甲基吡咯烷酮混合后,混合物可以直接进样检测分析,从而减少预处理的检测误差。加标回收率在92.21%~105.96%之间,验证了方法的准确性。此方法检测1,4-丁二醇中的金属元素,具有高效、干扰小、污染少、灵敏度高的优点。
Abstract: A rapid detection method for 39 trace metal elements in 1,4-butanediol was established. When using inductively coupled plasma mass spectrometry (ICP-MS) to detect organic substances, oxygen is added to reduce carbon deposition and avoid flameout. The temperature of the nebulization chamber should be set at 2˚C to provide a stable sample introduction environment. 1,4-butanediol is viscous with a freezing point of 20.1˚C. After mixing it with N-methylpyrrolidone, the mixture can be directly injected for detection and analysis, thus reducing the detection errors caused by sample pretreatment. The recoveries of the spiked samples range from 92.21% to 105.96%, which verifies the accuracy of the method. This method for detecting metal elements in 1,4-butanediol has the advantages of high efficiency, less interference, low pollution, and high sensitivity.
文章引用:李清霞, 梁斌, 张勇娟, 梁亚娟, 胜继伟. 1,4-丁二醇中的39种金属元素的检测方法[J]. 材料科学, 2025, 15(7): 1443-1451. https://doi.org/10.12677/ms.2025.157153

1. 引言

1,4-丁二醇(1,4-Butanediol,简写BDO),分子式为C4H10O2,分子量为90.12,外观为无色或淡黄色黏稠状液体,凝固点20.1℃。BDO是一种重要的精细化工基础原料,广泛应用于医药、化工、纺织、造纸、汽车和日用化学品等领域。BDO的下游产品有γ-丁内酯、四氢呋喃、聚氨酯树脂、增塑剂,并作为溶剂和电镀行业的增亮剂等[1]-[3]。随着1,4-丁二醇(BDO)的应用领域不断拓展,对其生产过程中金属元素的准确、高效定量分析需求日益增加。然而,目前关于BDO中金属元素分析检测方法的研究文献仍较为匮乏。

当前,金属元素的检测技术主要分为四类:原子吸收光谱(AAS) [4]、原子荧光光谱法(AFS)、电感耦合等离子体发射光谱法(ICP-OES)、电感耦合等离子体质谱法(ICP-MS) [5]。其中,原子吸收光谱法(AAS)通过使用与目标元素匹配的特征波长光线,使电子从基态跃迁至高能级状态。在此方法中,吸光度与目标元素的浓度呈正相关关系。然而,该方法的应用范围相对有限,在检测不同元素时需更换光源灯,且样品预处理步骤繁琐,检测限通常处于较大(≥1 mg/L)水平。原子荧光光谱法(AFS)易受到荧光猝灭效应和散射光干扰的影响,并且难以实现多元素的同时分析。电感耦合等离子体发射光谱法(ICP-OES)通常被认为是电感耦合等离子体质谱法(ICP-MS)的早期技术版本,其能够实现多元素的同时分析,尤其适用于低浓度金属元素的检测[6]。电感耦合等离子体质谱(ICP-MS)以其高效性、稳定性以及较少的化学干扰等优势,被广泛用于测定有机物中金属元素的定量分析[7]-[9]。随着科研人员的持续深入研究以及分析仪器技术的快速进步,ICP-MS (电感耦合等离子体质谱)已在半导体、合金、矿石、水资源、食品、日化产品及药品等多个领域中,被广泛用作检测金属元素含量的重要仪器[10]-[12]

目前,液体有机物的检测分析通常采用湿法消解、微波消解、稀释等方法去除有机质,随后利用ICP-MS进行金属元素的检测。然而,针对BDO中金属元素含量的测定,采用ICP-MS技术的相关报道较为有限。由于BDO中金属元素含量较低,加之繁琐的样品消解过程以及纯水稀释带来的干扰,可能导致测定结果误差较大。相比之下,使用有机溶剂进行样品稀释不仅可以有效控制污染来源,还能显著减少因水引入的测定误差。基于此,本研究选用有机溶剂NMP对BDO样品进行稀释处理,并直接进样通过ICP-MS进行检测分析[13] [14],并开展了检出限、精密度、加标回收率等方法验证。建立的此分析方法为BDO质量控制和应用提供科学依据。

2. 实验部分

2.1. 仪器与试剂

电感耦合等离子体质谱仪7800:美国安捷伦科技有限公司

制水机pro:德国赛多利斯

移液枪E4XLS:瑞士梅特勒托利多

枪头BioClean Ultra:瑞士梅特勒托利多

多元素混合标准溶液8500-6940 (10 μg/L):美国安捷伦科技有限公司

试剂瓶(PFA):日本亚速旺

天平(十万分之一):瑞士梅特勒托利多

硝酸(65%):德国默克

氩气(99.999%):濮阳市鑫源化工器材有限公司

氦气(99.999%):濮阳市鑫源化工器材有限公司

氩氧混合气(20%O, 80%Ar):濮阳市鑫源化工器材有限公司

N-甲基吡咯烷酮(99.92%):新迈奇材料股份有限公司

一四丁二醇(99.71%):陕西延长石油集团

2.2. 检测环境

因检测结果属于痕量级别,在普通的环境中容易对某些金属元素造成污染,避免引起检测误差,稀释样品、配置标准溶液、检测分析全程在百级洁净室内进行,并佩戴丁腈手套。

2.3. 加氧检测

ICP-MS分析有机样品时,雾化室需要冷却到2℃ [15],减少溶剂蒸汽进入等离子体,为气溶胶提供稳定的环境,保证气溶胶的均匀性,提高等离子体的稳定性和离子化效率,提高灵敏度和分析结果的稳定性。检测过程中,需要加入辅助气体氧气,减少锥孔积碳,避免熄火。加氧前后锥的积碳情况具体见表1

Table 1. The changes of the cone before and after oxygenation

1. 加氧前后的锥的变化

加氧前

加氧后

采样锥

截取锥

2.4. NMP稀释BDO

BDO为无色粘稠油状液体,凝固点20.1℃。在低温环境下直接进样时,BDO容易凝固,导致管路和雾化器堵塞。因此,本研究选择用N-甲基吡咯烷酮(NMP)对BDO稀释处理后进样。NMP的凝固点为−24.4℃,极性强,能与BDO完全互溶,有效降低BDO的黏度。BDO与NMP按照1:1比例混合后,在2℃时未出现凝固,可以正产进样并完成检测分析。

BDO用水稀释,因为所检样品中存在H2O,形成气溶胶的过程中可能会有少量的水进入到等离子体中心,会对某些金属离子造成干扰,会发生简单的化学反应,如35Cl和16O = 51ClO (51ClO = 51V信号叠加),水稀释后矾的检测值偏高,加标回收率约690%,造成矾(V)的检测误差,结果见表2

Table 2. Detection error of vitriol

2. 矾的检测误差

元素

水稀释加标前

水稀释加 1 μg/L标液

水稀释加标 回收率

NMP稀释 加标前

NMP稀释 加1 μg/L标液

NMP稀释 加标回收率

矾/V(μg/L)

0.2001

7.1773

697.7200

0.0258

1.0273

100.1500

0.1859

699.1400

0.0231

100.4200

0.2480

692.9300

0.0266

100.0700

2.5. 标准曲线的绘制

采用标准加入法,在BDO与NMP混合液(1:1)中加入多元素混合标准溶液,配制浓度为0.5 μg/L,1 μg/L,5 μg/L,10 μg/L,校准曲线由原始样品和各个加多元素标准溶液后样品的响应计数值(cps)与样品浓度绘图制得,未知样品的浓度通过反向延长校准曲线,得到X轴截距而得到未知样品浓度。

2.6. 仪器工作条件

利用ICP-MS检测有机物时,若采用直接进样方法,则需配置有机进样系统及氧气引入装置。通过氧气与有机物中的碳元素反应生成二氧化碳并排出,可有效减少碳元素对检测结果的干扰误差。同时,采用干燥纯净的He气碰撞模式,能够去除潜在干扰物,从而显著改善Cr52、Fe56、Cu63和As75的测量精度。此外,使用1.0 mm内径的中心管可减少有机物挥发物进入等离子体的可能性;选用铂锥,降低锥体损耗;PFA雾化器以及耐有机溶剂的管路系统。具体工作参数见表3

Table 3. Working parameters of ICP-MS

3. 电感耦合等离子体质谱仪工作参数

仪器参数

设定值

仪器参数

设定值

射频功率/W

1550W

采样深度/mm

8

载气Ar流量/(L∙min1)

0.5

同心雾化器材质

PFA

可选气Ar流量/(L∙min1)

15

炬管材质

石英

碰撞气体He流量/(L∙min1)

2

采样锥材质

Pt

辅助气ArO流量/(L∙min1)

0.2

截取锥材质

Pt

冲洗泵速/(r∙min1)

0.5

透镜电压/V

10

分析泵速/(r∙min1)

0.1

偏转电压/V

−90

样品引入时间(s)

30

3. 结果与讨论

3.1. 线性关系和检出限

在100 mL全氟烷氧基树脂(PFA)瓶中加1:1的BDO与NMP的混合液作为0 μg/L,之后加入混合标准溶液,分别浓度为0.5 μg/L,1 μg/L,5 μg/L,10 μg/L,以各元素浓度为横坐标,以待测元素cps计数值为纵坐标,做标准曲线,结果显示39种金属元素在其浓度范围内线性关系良好,具体结果见表4表5

Table 4. The cps counts of each element at different concentrations

4. 各元素在不同浓度的cps计数值

cps计数值

0 μg/L

0.5 μg/L

1 μg/L

5 μg/L

10 μg/L

锂/Li

54.45

3590.47

7107.37

36508.14

73806.45

铍/Be

2.22

1300.07

2841.41

14045.63

28510.78

硼/B

790.03

1541.21

2389.11

9956.77

20324.03

钠/Na

7738.77

13236.07

18722.06

59294.47

112526.64

镁/Mg

2.22

125.56

254.45

1597.88

3107.03

铝/Al

3.33

50.00

106.67

635.58

1134.50

钾/K

61.11

122.23

261.12

1393.41

2602.48

钙/Ca

60.00

192.23

316.68

1427.86

2513.58

钛/Ti

727.81

1200.06

1584.55

5630.04

10490.57

钒/V

601.86

6270.33

12822.81

63471.03

129305.34

铬/Cr

108.89

334.45

554.46

2220.20

4399.61

锰/Mn

268.95

4674.23

9086.38

44224.58

87189.87

铁/Fe

196.67

1987.94

6819.02

18466.40

36575.42

钴/Co

10.00

6045.78

12784.70

65033.26

132163.10

镍/Ni

81.11

1004.49

1903.54

8753.86

17339.54

铜/Cu

327.51

3439.34

7335.07

36346.52

73285.34

锌/Zn

17.78

637.80

1222.29

5943.52

11758.28

锗/Ge

28.89

1397.91

2852.55

13830.18

27760.48

砷/As

3774.98

6034.89

7795.56

24290.91

45199.45

铷/Rb

112.22

6035.81

12194.26

62082.54

125782.94

锶/Sr

2.22

8215.83

16932.50

87840.53

176610.17

锆/Zr

1.11

5615.62

11456.99

57072.99

115713.72

铌/Nb

4.44

10720.84

21359.70

106444.49

213400.32

铑/Rh

2.22

11691.67

24147.70

121148.74

245911.07

银/Ag

1.11

3826.13

8046.87

40405.29

82563.28

钯/Pd

3.33

3092.61

6379.33

32122.73

64596.00

镉/Cd

0

808.92

1600.12

8418.19

16435.39

铟/In

4.44

12941.72

26347.40

132260.87

267808.46

锡/Sn

5.55

2664.73

5440.03

28219.98

56698.05

锑/Sb

1.11

4071.76

8532.72

42802.98

87885.82

铯/Cs

103.33

11685.11

23406.82

119172.01

241654.28

钡/Ba

31.11

1503.43

3037.05

14969.44

30589.96

钽/Ta

3.33

21129.12

43186.75

215583.33

434993.08

钨/W

2.22

5737.97

11452.82

57344.96

115507.16

铂/Pt

6.66

2466.93

5135.49

25100.72

50883.62

金/Au

0

4354.10

8905.32

44914.27

88914.48

铊/Tl

5.56

13716.18

27956.66

139275.38

280939.26

铅/Pb

5.67

9638.23

19693.12

98244.13

198700.15

铋/Bi

26.67

16824.30

33963.09

168463.70

337456.57

用PFA瓶中的BDO与NMP混合液,做空白,39种元素扫描11次,以仪器工作站自动算出的标准偏差,根据方法检出限计算方法,计算出方法检出限[16],具体结果见表5

检出限(MDL)计算公式为:

MDL = 3S/k

式中,S为标准偏差,k为斜率

Table 5. Standard curve relationship, detection limit and precision

5. 标准曲线关系、检出限、精密度

元素

线性方程

相关系数/r

方法检出限/(μg/L)

精密度RSD/%

锂/Li

Y = 7355.3218x + 54.4467

1

0.0001

0.1661

铍/Be

Y = 2841.9071x + 2.2200

1

0.0001

0.0340

硼/B

Y = 1925.9271x + 790.0333

0.9995

0.0001

0.4355

钠/Na

Y = 10450.5985x + 7738.7733

0.9999

0.0001

0.2452

镁/Mg

Y = 311.6281x + 2.2200

0.9997

0.0001

0.0012

铝/Al

Y = 196.7623x + 376.6767

0.9992

0.0006

0.0002

钾/K

Y = 256.0895x + 61.1133

0.9994

0.0003

0.2152

钙/Ca

Y = 245.6153x + 60.0000

1

0.0003

0.0002

钛/Ti

Y = 976.0927x + 727.8067

0.9999

0.0002

2.8844

钒/V

Y = 12803.9282x + 601.8567

0.9999

0.0001

0.1932

铬/Cr

Y = 427.8974x + 108.8933

0.9999

0.0003

0.1381

锰/Mn

Y = 8712.9306x + 268.9467

1

0.0001

0.6018

铁/Fe

Y = 3640.9725x + 196.6700

1

0.0001

0.1746

钴/Co

Y = 13167.8405x + 10.0000

1

0.0001

0.0179

镍/Ni

Y = 1728.5712x + 81.1133

1

0.0002

0.4391

铜/Cu

Y = 7273.1630x + 327.5100

1

0.0001

0.0665

锌/Zn

Y = 1176.6195x + 17.7800

1

0.0001

0.1606

锗/Ge

Y = 2770.9353x + 28.8867

1

0.0001

0.0685

砷/As

Y = 4134.4540x + 3774.9833

1

0.0001

1.6113

铷/Rb

Y = 12527.5454x + 112.2233

1

0.0001

0.0544

锶/Sr

Y = 17634.1238x + 2.2233

1

0.0001

0.0382

锆/Zr

Y = 11538.6073x + 1.1100

1

0.0001

0.0082

铌/Nb

Y = 21329.2865x + 4.4433

1

0.0001

0.0112

铑/Rh

Y = 24513.3561x + 2.2233

1

0.0001

0.0075

银/Ag

Y = 8218.6207x + 1.1100

1

0.0001

0.0879

钯/Pd

Y = 6451.0438x + 3.3333

1

0.0001

0.0129

镉/Cd

Y = 1651.0848x + 0.0000E + 000

0.9999

0.0001

0.0014

铟/In

Y = 26709.9717x + 4.4433

1

0.0001

0.0120

锡/Sn

Y = 5661.4749x + 5.5533

1

0.0001

0.0289

锑/Sb

Y = 8739.9876x + 1.1100

0.9999

0.0001

0.0014

铯/Cs

Y = 24078.7900x + 103.3333

1

0.0001

0.0206

钡/Ba

Y = 3041.7600x + 31.1100

0.9999

0.0001

0.0149

钽/Ta

Y = 43418.1690x + 3.3333

1

0.0001

0.0049

钨/W

Y = 11513.3195x + 2.2200

1

0.0001

0.0067

铂/Pt

Y = 5074.0501x + 6.6633

1

0.0001

0.0186

金/Au

Y = 8980.5827x + 0.0000E + 000

1

0.0001

0.0051

铊/Tl

Y = 28043.5055x + 5.5567

1

0.0001

0.0189

铅/Pb

Y = 19822.8900x + 5.6677

1

0.0001

0.0111

铋/Bi

Y = 33733.2235x + 26.6677

1

0.0001

0.0047

由以上结果,各元素线性关系良好,相关系数在0.9990~1.0000之间,检出限小于0.001 μg/L。复测后,检测数据显示仪器精密度RSD值小于5。根据GB/T39486-2020电感耦合等离子体质谱分析方法通则要求,所检元素含量小于1 μg/L时,精密度RSD值小于15,所选方法满足检测要求。

3.2. 加标回收率

将各金属元素浓度为1 μg/L混合标准溶液液加入在装有BDO与NMP的混合液的PFA瓶中,将各元素扫描11次检测所得值,做加标回收率,结果见表6

Table 6. The standard recovery rate of each element

6. 各元素的加标回收率

测定元素

本底值/(μg/L)

加标量/(μg/L)

测定值/(μg/L)

回收率/%

锂/Li

0.0103

1

1.0290

101.87

铍/Be

0.0012

1

1.0221

102.09

硼/B

0.2515

1

1.2688

101.73

钠/Na

0.8888

1

1.9041

101.53

镁/Mg

0.0143

1

0.9520

93.77

铝/Al

0.0192

1

0.9413

92.21

钾/K

0.2213

1

1.1628

94.15

钙/Ca

0.2124

1

1.1459

93.35

钛/Ti

0.8390

1

1.7827

94.37

钒/V

0.0258

1

1.0273

100.15

铬/Cr

0.2727

1

1.2101

93.74

锰/Mn

0.0295

1

0.9669

93.74

铁/Fe

0.0543

1

1.0935

103.92

钴/Co

0.0015

1

1.0085

100.7

镍/Ni

0.0681

1

1.0260

95.79

铜/Cu

0.0136

1

1.0605

104.69

锌/Zn

0.0349

1

1.0945

105.96

锗/Ge

0.0136

1

1.0648

105.12

砷/As

0.9746

1

1.9621

98.75

铷/Rb

0.0085

1

1.0191

101.06

锶/Sr

0.0002

1

0.9913

99.11

锆/Zr

0.0023

1

1.0054

100.31

铌/Nb

0.0002

1

1.0054

100.52

铑/Rh

0.0004

1

0.9941

99.37

银/Ag

0.0019

1

0.9810

97.91

钯/Pd

0.0005

1

1.0139

101.34

镉/Cd

0.0007

1

1.0223

102.16

铟/In

0.0010

1

1.0248

102.38

锡/Sn

0.0031

1

1.0080

100.49

锑/Sb

0.0013

1

0.9970

99.57

铯/Cs

0.0036

1

1.0062

100.26

钡/Ba

0.0073

1

1.0196

101.23

钽/Ta

0.0001

1

1.0281

102.8

钨/W

0.0008

1

1.0230

102.22

铂/Pt

0.0004

1

1.0222

102.18

金/Au

0.0006

1

1.0337

103.31

铊/Tl

0.0011

1

1.0025

100.14

铅/Pb

0.0006

1

1.0135

101.29

铋/Bi

0.0006

1

1.0239

102.33

由以上结果看出,加标回收率在92.21%~105.96%,符合GB39486-2020中待测元素含量对应的正确度范围。

4. 结论

BDO与NMP混和后,凝固点升高了,采用ICP-MS直接进样,测定1,4-丁二醇中的39种金属元素。线性相关系数在0.9990~1.0000之间,加标回收率92.21%~105.96%,结果表明,此检测方法高效、误差小、干扰小,此方法适用于凝固点低的有机物中金属元素的检测。

NOTES

*共同第一作者。

#通讯作者。

参考文献

[1] 魏国玉, 侯艺琳. Reppe法1,4-丁二醇铜铋催化剂国产应用[J]. 上海化工, 2024, 49(4): 29-32.
[2] 吴慧玲, 孟令臣, 张思琦, 等. 1, 4-丁二醇制备γ-丁内酯脱氢工艺[J]. 石油化工, 2025, 54(3): 297-303.
[3] Raju, M.A., Gidyonu, P., Nagaiah, P., Rao, M.V., Raju, B.D. and Rao, K.S.R. (2019) Mesoporous Silica-Supported Copper Catalysts for Dehydrogenation of Biomass-Derived 1,4-Butanediol to Gamma Butyrolactone in a Continuous Process at Atmospheric Pressure. Biomass Conversion and Biorefinery, 9, 719-726.
https://doi.org/10.1007/s13399-019-00406-4
[4] 杨虹. 原子吸收光谱法在测定水中重金属的应用研究[J]. 山西化工, 2021, 41(5): 74-77.
[5] 冯先进, 郑国经. 电感耦合等离子体发射光谱与质谱分析技术和应用研究进展[J]. 中国无机分析化学, 2025, 15(3): 353-362.
[6] Mdluli, N.S., Nomngongo, P.N. and Mketo, N. (2024) Ionic Liquid Assisted Extraction Induced by Emulsion Breaking for Extraction of Trace Metals in Diesel, Gasoline and Kerosene Prior to ICP-OES Analysis. Heliyon, 10, e26605.
https://doi.org/10.1016/j.heliyon.2024.e26605
[7] Hirata, T., Kobayashi, K., Khoo, H.H., Shikino, O. and Asanuma, H. (2024) Detection of Several Volatile Organic Compounds through Ar+ Induced Chemical Ionisation Using Inductively Coupled Plasma-Tandem Mass Spectrometry (ICP-MS/MS). The Analyst, 149, 5174-5183.
https://doi.org/10.1039/d4an00996g
[8] 汤地生, 杨婉晴, 张宇, 等. 电感耦合等离子体质谱(ICP-MS)法测定高钨基体样品中稀土元素的含量[J]. 中国无机分析化学, 2024, 14(4): 457-463.
[9] 黄丹宇. 电感耦合等离子体质谱法测定镁合金中的痕量元素[J]. 分析测试技术与仪器, 2023, 29(3): 328-333.
[10] Lossow, K., Schlörmann, W., Tuchtenhagen, M., Schwarz, M., Schwerdtle, T. and Kipp, A.P. (2023) Measurement of Trace Elements in Murine Liver Tissue Samples: Comparison between ICP-MS/MS and TXRF. Journal of Trace Elements in Medicine and Biology, 78, Article 127167.
https://doi.org/10.1016/j.jtemb.2023.127167
[11] Sim, K.S., Kim, H., Hur, S.H., Na, T.W., Lee, J.H. and Kim, H.J. (2024) Geographical Origin Discriminatory Analysis of Onions: Chemometrics Methods Applied to ICP-OES and ICP-MS Analysis. Food Research International, 175, Article 113676.
https://doi.org/10.1016/j.foodres.2023.113676
[12] 万曦, 田海花, 刘涌, 等. ICP-MS测定电子级丙二醇甲醚中的钠、镁、钙、铁和铅含量[J]. 2023, 42(1): 90-94.
[13] 张文媚, 吕玉平, 闻环, 等. 有机溶剂稀释-电感耦合等离子体发射质谱法测定原油砷含量[J]. 广东化工, 2022, 49(3): 179-181.
[14] 李丹. 基于有机溶剂前处理的ICP-MS法直接测定盐酸溴己新中的汞含量[J]. 广州化工, 2025, 53(8): 160-163.
[15] 张磊, 唐辉, 王辉, 等. 电感耦合等离子体质谱法直接测定有机相中痕量锆[J]. 质谱学报, 2015, 36(2): 156-160.
[16] 李春华, 陈鹰, 韩宝英, 等. 化学试剂 电感耦合等离子体质谱分析方法通则: GB/T 39486-2020 [S]. 2020.