提高SiPM时间分辨能力方法研究——基于两种SiPM性能对比测试
Study on the Method of Improving Time Resolution of SiPM—Based on the Testing of Two Types of SiPMs Coupled with YSO
DOI: 10.12677/OJCS.2019.82003, PDF,    科研立项经费支持
作者: 李维姣, 陈嘉敏:公安部第三研究所,上海;陈 瑞, 奚道明:苏州瑞迈斯医疗科技有限公司,江苏 苏州
关键词: SiPMYSO晶体能量分辨率零极相消电路 SiPM YSO Crystal Energy Resolution Pole-Zero Circuit
摘要: 本文通过实验分析从脉冲宽度和能量分辨率两个维度,对比Sensl的FC10035和FC10020的性能,为基于硅光电倍增管(SiPM)的闪烁体探测器提供设计参考,实验表明FC10020的脉冲宽度仅为200 nS,适合对于时间性能要求较高的应用场景,FC10035的能量分辨能力较好,适用于对能量分辨要求较高的应用场景;为提高响应的动态范围,降低Pileup的概率,设计了pole-zero (零极相消)电路,通过实验验证了零极相消电路对于缩短脉冲宽度的有效性。
Abstract: This paper compares the performances of Sensl’s FC10035 and FC10020 from the two dimensions of pulse width and energy resolution, and provides a reference for the design of scintillator detector based on SiPM. Experiments show that the pulse width of FC10020 is only 200 nS, which is suitable for application scenarios with high time performance requirements. The energy resolution of FC10035 is suitable for application scenarios of high energy performance requirements. In order to improve the dynamic range of response and reduce the probability of pile up, a pole-zero cancellation circuit is designed, and the effectiveness of the zero-pole cancellation circuit for shortening the pulse width is verified by experiments.
文章引用:李维姣, 陈嘉敏, 陈瑞, 奚道明. 提高SiPM时间分辨能力方法研究——基于两种SiPM性能对比测试[J]. 电路与系统, 2019, 8(2): 16-22. https://doi.org/10.12677/OJCS.2019.82003

参考文献

[1] Risigo, F., Bulgheroni, A., Caccia, M., et al. (2009) SiPM Technology Applied to Radiation Sensor Development. Nu-clear Instruments & Methods in Physics Research, Section A (Accelerators, Spectrometers, Detectors and Associated Equipment), 607, 75-77. [Google Scholar] [CrossRef
[2] Shi, F., Liu, J.G., Hong, L.U., et al. (2011) Comparative Studies of Silicon Photomultipliers and Traditional Vacuum Photomultiplier Tubes. Chinese Physics C: English Edition, 35, 50-55. [Google Scholar] [CrossRef
[3] 赵景泰, 王红, 金滕滕, 等. 闪烁晶体材料的研究进展[J]. 中国材料进展, 2010, 29(10): 40-48.
[4] Otte, A.N., Garcia, D., Nguyen, T., et al. (2017) Characterization of Three High Efficiency and Blue Sensitive Silicon Photomultipliers. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 846, 106-125. [Google Scholar] [CrossRef
[5] Gola, A., Piemonte, C. and Tarolli, A. (2013) Analog Circuit for Timing Measurements with Large Area SiPMs Coupled to LYSO Crystals. IEEE Transactions on Nuclear Science, 60, 1296-1302. [Google Scholar] [CrossRef