稀有气体分析仪的校准方法研究及其不确定度评定示例
Calibration Research of Noble Gas Analyser and Its Uncertainty Evaluation Example
DOI: 10.12677/IaE.2022.102009, PDF,    科研立项经费支持
作者: 黄梓宸, 谢昭群, 郭 威, 许俊斌, 黄彦捷:广东省计量科学研究院,广东 广州
关键词: 稀有气体分析仪校准不确定度Noble Gas Analyser Calibration Uncertainty
摘要: 稀有气体分析仪是一类用于分析特定环境下稀有气体浓度的仪器。基于稀有气体分析仪的原理及应用,提出了校准过程中应满足的试验条件及主要技术指标,开发了示值误差、重复性、响应时间和报警功能的校准方法,并选取典型样品进行了试验论证,对测量过程的不确定度进行了分析。结果表明所提出的指标及校准方法切实可行。
Abstract: Noble gas analyser is a type of instrument used to analyze the concentration of noble gas in specific environment. The requirements for experimental condition and technical indicators are made based on the principle and application. Also, calibration methods for items including error of indication, repeatability, response time and alarm function are proposed, typical sample is selected for experimental demonstration, and the uncertainty of the measurement process is analyzed. The results show that the proposed index and calibration method are feasible.
文章引用:黄梓宸, 谢昭群, 郭威, 许俊斌, 黄彦捷. 稀有气体分析仪的校准方法研究及其不确定度评定示例[J]. 仪器与设备, 2022, 10(2): 66-72. https://doi.org/10.12677/IaE.2022.102009

参考文献

[1] 梁爱琴, 杨曼丽, 姚文红. 稀有气体及其化合物的发现和应用[J]. 化学教学, 2007(2): 47-49
[2] 谢昭群, 张猛, 贾钠钧, 等. 环氧乙烷检测仪的校准方法研究[J]. 仪器与设备, 2019, 7(1): 53-57. [Google Scholar] [CrossRef
[3] 郭威, 黄彦捷, 花秀兵. 碳平衡法汽车燃料消耗量检测仪校准方法开发[J]. 仪器与设备, 2021, 9(2): 29-34. [Google Scholar] [CrossRef
[4] 上海市技术监督局. JJG663-90. 热导式氢分析器[S]. 上海. 国家技术监督局, 1990.
[5] 全国环境化学计量技术委员会. JJG635-2011. 一氧化碳、二氧化碳红外气体分析器检定规程[S]. 北京: 国家质量监督检验检疫总局, 2011.
[6] 陈明华. 测量仪器准确度、最大允许误差和不确定度辨析[J]. 中国计量, 2003(6): 67-68.
[7] 全国法制计量技术委员会. JJF1059.1-2012. 测量不确定度评定与表示[S]. 北京: 国家质量监督检验检疫总局, 2012.
[8] 全国法制计量技术委员会. JJF1094-2002. 测量仪器特性评定[S]. 北京: 国家质量监督检验检疫总局, 2002.